James William Mckeage, Kieran Brennan1, Geehoon Park, N. Catherine Hogan, Ian Hunter, Bryan Ruddy2, Poul Nielsen, Andrew Taberner2
14:50 - 15:05 | Wed 12 Jul | Dunn Room | WeBT13.3
The effect of varying velocity during jet injection on the dispersion of fluid into tissue is investigated using a custom-built X-ray imaging system. Injections are performed into ex-vivo porcine abdominal tissue using a voice coil actuated injection device. Single velocity and two-phase velocity injections reveal the complex nature of the dispersion of the fluid jet in layered tissue and highlight the effects of changing the jet velocity following the initial penetration of the liquid into the tissue.