Origami Microfluidic Paper-Analytical-Devices (omPAD) for Sensing and Diagnostics

Details

09:00 - 09:15 | Wed 17 Aug | Sorcerers Apprentice 3 | WeAT15.5

Session: Micro/Nanofabricated Devices for Biomedical Sensing and Diagnostics

Abstract

Recent research activities in the area of low-cost sensing and diagnostics that are realized on cellulosic paper substrate are presented. First a three-dimensional origami paper-based analytical device (omPAD) with multiple electrochemical sensors, an integrated sample reservoir and tight integration with a custom CMOS potentiostat is presented. Second, an optical sensor array with built-in microfluidic channel for sample delivery is presented. The sensors are fabricated using a combination of wax printing and screen-printing using a solution based approach in ambient conditions without the need for expensive fabrication equipment or a cleanroom. Readout is based on using existing consumer grade electronic devices like flatbed scanner (for optical sensor) or custom designed CMOS potentiostat (for electrochemical sensors). Together the 3D paper-based analytical device with integrated sensor, microfluidics and portable readout instrumentation demonstrates a low-cost, self-contained system suitable for sensing and point-of-care diagnostics.